Common product failures and solutions
Fault Description | Possible Reason | Solutions |
Unable to Power On | Low battery | Replace new batteries |
The battery is not installed | Install the battery correctly according to the battery label | |
Circuit fault | Contact dealers or manufacturer | |
No response when blowing | Not fully warmed up | Warm up several times |
Circuit fault | Contact dealers or manufacturer | |
No results are displayed when blowing | Residual gas in the product | Rewarm up the product |
Not blowing within the time | Please blow during "BIOW" | |
Have results without blowing | Residual gas in the product | Warm up several times |
Several test results are different | Test immediately after drinking | Test 20 minutes after drinking |
The blowing time is different | It is recommended to take a deep breath for five seconds | |
Wrong Display | Expired sensor | Contact dealers or manufacturer to replace the Sensor |
Over the specified working conditions | Please be used in specified conditions | |
Long time no calibration | Calibrate the tester in time |